Abutment Screw Failure Caused by Off-angle Loading on an Inclined Maxillary Implant – 0043
This case showcases an example of Abutment screw failure resulting from off axis loading on a lingually inclined maxillary implant.
This case showcases an example of Abutment screw failure resulting from off axis loading on a lingually inclined maxillary implant.
This implant case showcases an example of a Recovery of a Fractured Abutment and Abutment Screw from a Straumann RC BL Implant.
This implant case exemplifies Ultrasonic thread damage during abutment screw retrieval secondary to NC components in an RC Straumann implant.
This case exemplifies a fractured abutment screw case initially attempted with a generic recovery kit and no microscopic visualization.
This case illustrates an example of a Mechanical Assessment of a Straumann RC BL Case Where Two Abutments had Failed in Two Years.
This case illustrates how wrong restorative components led to two abutment screw failures in two years leading to an abutment screw rescue.
The patient presented for retrieval of a fractured healing abutment screw from a Straumann 4.1 RC BL implant in the #13 site
Latrogenic Inclusion of Impression Material Resulting in an Early Abutment Screw Fracture. The patient presented for retrieval of a fractured abutment screw from a Straumann 4.8 BL implant in the #19 site.
The patient presented on referral for retrieval of a fractured abutment from a 4.8 Straumann BL RC implant in the #18 site.
The patient presented for recovery of an abutment and abutment screw from a Straumann 4.8 BL RC implant in the #30 site.